Measure the whole impedance,
not just the face your DMM shows you.
The DER EE DE-5000 dual-display LCR meter from first power-on to calibrated, tolerance-sorted measurements: the impedance math, the panel, open/short calibration, series vs parallel, and reading D/Q/ESR/θ like you mean it.
1 · Mental model A DMM tells you a number. This tells you a vector.
A multimeter's capacitance function gives you one number and lets you assume the part is ideal. An LCR meter refuses that assumption. It measures impedance — a quantity with two parts, a resistive part and a reactive part — and only then computes the "capacitance" or "inductance" you asked for, plus a second number that tells you how much the real part deviates from the ideal one. Every button on the DE-5000 exists to choose which face of that vector you see.
Level 1 · why "capacitance" alone is an incomplete answerskip if fluent
Impedance has a real part and an imaginary part
In series form, impedance is Zs = Rs + jXs: a resistance Rs (the real part, the part that dissipates energy as heat) plus a reactance Xs (the imaginary part, the part that stores and returns energy). The magnitude is |Z| = √(Rs² + Xs²) and the phase angle is θ = tan⁻¹(Xs/Rs). A DMM's capacitance mode reports a single scalar and throws away everything about Rs. The DE-5000 keeps it.
Reactance has a sign, and the sign tells you what kind of part you have
Two kinds of reactance exist: inductive reactance XL = 2πfL and capacitive reactance XC = 1/(2πfC), where f is the test signal's frequency. The manual's convention: if θ > 0 the reactance is inductive; if θ < 0 it's capacitive. This is why the same auto-detect logic that picks L vs C vs R (§5) is really just reading the sign and magnitude of one angle.
A "pure" component doesn't exist off the datasheet
Every real capacitor has some series resistance (from its leads, plates and dielectric losses) and every real inductor has some parallel leakage. The ratio between the reactive part and the resistive part — quality factor Q for inductors, dissipation factor D for capacitors, where D = 1/Q = tanθ — is the number that tells you how close to ideal the part actually is. A 100 µF electrolytic with a bad D is a worse capacitor than its printed value suggests, and no DMM will ever tell you that.
Series and parallel are two descriptions of the same impedance, not two different DUTs
The same physical component can be described as a resistance and a reactance in series (Zs = Rs + jXs) or as a conductance and a susceptance in parallel (Y = 1/Z = G + jB, admittance). Both describe the identical impedance; they're duals. Which one is the more useful description depends on whether the loss mechanism in your specific part behaves more like a series resistor or a parallel leakage path — the subject of §7.
1.1 · The impedance vector plane
The manual draws this exact picture to introduce the whole instrument. Resistance Rs runs along the real axis; reactance Xs runs along the imaginary axis. A part with positive reactance (inductive) plots above the real axis at a positive angle θ; a part with negative reactance (capacitive) plots below it at a negative angle θ₁. The vector's length is |Zs|, its angle is θ, and every measurement the DE-5000 reports is a projection of this one picture onto a display.
Sources: manual §II Measuring Principle Introduction, pp.4–7 (Impedance parameter introduction, Measurement mode, Open/short calibration).
2 · Prerequisite floor What this guide assumes, and what it doesn't
You don't need a physics degree to run this meter well. You need three ideas solid enough that you can lean on them without re-deriving them, and this guide will name every place it uses them.
2.1 · Self-check
- I can use
V = IRand know what "resistance" physically resists. - I know that reactance (from a capacitor or inductor) opposes current in a way that depends on frequency, and that it stores energy rather than burning it as heat.
- I can tell a series circuit from a parallel circuit by eye: series shares one current path, parallel splits current across branches.
- I'm comfortable with engineering notation — p / n / µ / m / k / M — as it appears on a capacitor or inductor's printed value.
Sources: manual §II (pp.4–7) supplies the formulas this floor leans on.
3 · Panel & display orientation Battery in, power on, and know what you're looking at
Fourteen keys, one dual-line LCD, and four input terminals. Before you measure anything, learn where the keys are and what the twenty-nine icons around the digits mean — the rest of this guide refers back to this map constantly.
3.1 · Battery and first power-on
The DE-5000 ships with a standard accessory set: the TL-21 alligator test lead case, an AC/DC adaptor, the TL-23 guard line, the user manual, and a 9 V battery (NEDA 1604 / JIS006P / IEC6F22 carbon-zinc or alkaline — alkaline recommended). To install or replace it: remove the tilt-stand, loosen the four screws on the battery cover with a suitable screwdriver, lift the cover, fit the battery observing polarity, and reverse the procedure.
- Press POWER. Expect: every LCD segment and icon lights for two seconds — the manual's own display self-test — then the meter settles into its default state.
- Read the default state. Expect: LCR Auto mode, test frequency 1 kHz. This is the meter's power-on default every time, not a mode you have to select.
OFF and powers down. APO is enabled — and the APO icon shows — only when running on the 9 V battery. Run the meter from the AC/DC adaptor and APO disables itself automatically; the APO icon disappears.
3.2 · Front panel
Every key does one job in LCR Auto mode (measurement primitives) and often a second job once you leave it (SORTING/SETUP navigation). The diagram below follows the manual's own panel illustration and numbering.
| # | Item | What it does |
|---|---|---|
| 1 | LCD display | Dual 19999 / 9999-count display, plus icon row |
| 2 | POWER | Power the instrument on/off |
| 3 | LCR AUTO | Cycles Auto-LCR → Auto-L → Auto-C → Auto-R → Auto-DCR (§5) |
| 4 | FREQ | Cycles test frequency: 1 kHz → 10 kHz → 100 kHz → 100 Hz → 120 Hz (§5) |
| 5 | ☀ (backlight) | Toggles the LCD backlight; auto-off after 60 s idle |
| 6 | SORTING | Enters tolerance-binning mode (§9); disabled in LCR Auto mode |
| 7 | PC | Starts/stops UART data transmission to the optional IR-to-USB adapter (§10) |
| 8 | CAL | Open/short calibration (§4) |
| 9 | D/Q/ESR/θ | Cycles the secondary-display parameter (§6); disabled in LCR Auto mode |
| 10 | SETUP | Setup-menu navigation inside SORTING mode (§9) |
| 11 | SER/PAL | Manually forces series or parallel mode (§7); disabled in LCR Auto mode |
| 12 | ENTER | Confirms a SORTING setup step (§9) |
| 13 | REL% | Relative (delta) mode (§8); disabled in LCR Auto mode |
| 14 | HOLD | Freezes the primary display (§8) |
| 15 | Input sockets and terminals | GUARD, −, + — 4-wire sockets plus guard |
| 16 | AC/DC power adaptor jack | Side panel; disables APO automatically when in use |
| 17 | Battery cover | Rear; four screws |
| 18 | Tilt-stand | Rear |
| 19 | IR-to-USB slot | Rear; mates with the optional IR-to-USB case |
| 20 | IR to USB case (optional) | Not included — see §10 |
| 21 | TL-21 alligator test lead case | Standard accessory; the leads used throughout this guide |
| 22 | TL-22 SMD tweezers (optional) | Not included — mentioned where it changes a lab |
| 23 | TL-23 guard line | Standard accessory; shields the DUT from interference at high impedance (§4) |
3.3 · LCD icon map
The DE-5000's LCD carries thirty numbered icons around the two rows of digits. You don't need to memorize all of them — you'll absorb the ones you use constantly (Auto, LCR, the mode letters) in the first hour — but this table is the ground truth to check against when an icon you don't recognize appears.
| # | Icon | Meaning |
|---|---|---|
| 1 | Sorting | Sorting function is enabled |
| 2 | Tol | Tolerance in sorting mode: ±0.25%, ±0.5%, ±1%, ±2%, ±5%, ±10%, ±20%, or +80%/−20% |
| 3 | kHz | Test frequency: 1 kHz, 10 kHz, 100 kHz, 100 Hz, or 120 Hz |
| 4 | PC | Communication is active |
| 5 | battery icon | Battery capacity indication |
| 6 | Range | Range selection enabled, on the sorting setup menu |
| 7 | Auto | Auto range for L, C, or R measurement |
| 8 | LCR | Checking for L/C/R mode automatically |
| 9 | Δ | Relative function (REL%) is enabled |
| 10 | Ls/Lp | Inductance in series or parallel mode is active |
| 11 | Cs/Cp | Capacitance in series or parallel mode is active |
| 12 | Rs/Rp | AC resistance in series or parallel mode is active |
| 13 | DCR | DC resistance mode is selected |
| 14 | D/Q/θ | Dissipation factor, quality factor, or phase angle — secondary L/C parameter |
| 15 | Rp | AC resistance in parallel mode is active |
| 16 | Cal | Open/short calibration mode |
| 17 | HOLD | Data hold is active |
| 18 | APO | Auto power-off mode (battery power only) |
| 19 | ESR | Series equivalent resistance mode |
| 20 | secondary display | The smaller readout |
| 21 | ° | Phase-angle unit |
| 22 | MkΩ (secondary) | Unit for resistance (Ω, kΩ, MΩ) on the secondary display |
| 23 | pnµF (secondary) | Unit for capacitance (pF, nF, µF, mF) on the secondary display |
| 24 | µmH (secondary) | Unit for inductance (µH, mH, H) on the secondary display |
| 25 | % | Percentage display in relative mode, on the secondary display |
| 26 | primary display | The larger readout |
| 27 | MkΩ (primary) | Unit for resistance on the primary display |
| 28 | µmH (primary) | Unit for inductance on the primary display |
| 29 | pnµF (primary) | Unit for capacitance on the primary display |
| 30 | bar-graph | Analogue bar-graph display, 0–100% |
Two special indication strings also appear in place of a number: OPEn during open calibration and Srt during short calibration (§4).
Sources: manual §III Panel Illustration (pp.8–10) and §III LCD Display Illustration (pp.11–12); §V Replacing Batteries (p.33).
4 · Calibration Open/short: cancelling the fixture so you measure the part, not the clips
Every measurement passes through a test fixture — sockets, leads, clips — and that fixture is not impedance-free. It adds a small series impedance and a small parallel admittance of its own, and at the extremes of the DE-5000's range those parasitics are not small relative to what you're trying to measure. CAL doesn't calibrate the DUT; it calibrates the fixture out of the picture.
4.1 · Why the fixture pollutes the reading
The manual models the measured impedance ZM as the true DUT impedance ZDUT in series with a fixture's series impedance, in parallel combination with the fixture's parallel admittance:
ZM = (Rs + jωLs) + [(Go + jωCo)⁻¹ ∥ ZDUT]
The open measurement captures the fixture's parallel admittance in isolation — with nothing connected, the current through ZDUT is zero and Y_OPEN = Go + jωCo is all that's left to measure. The short measurement captures the fixture's series impedance in isolation — with the leads shorted, Z_SHORT = Rs + jωLs is all that's left. With both known, the meter solves for the true DUT impedance on every subsequent measurement:
ZDUT = (ZM − Z_SHORT) / [1 − (ZM − Z_SHORT) · Y_OPEN]
4.2 · Running the calibration
FAIL. Use the TL-21 leads (or an improvised lead — but not a long one) exactly as you intend to measure with afterward, since the calibration is specific to that fixture.
- Connect TL-21 to the meter's terminals, leaving the far end open. Hold CAL for 2 seconds. Expect: the display enters open-calibration mode, showing
OPEnwith a blinking----. - Press CAL to start. Expect: a 30-second countdown runs (
OPEn -30-counting down), thenOPEn PASSorOPEn FAIL. - Press CAL again. Now short the TL-21 leads together. Expect: the display enters short-calibration mode,
Srt ----. - Press CAL to start. Expect: another 30-second countdown (
Srt -30-), thenSrt PASSorSrt FAIL. - If both steps show
PASS, press CAL one more time to store the calibration to EEPROM and return to measurement mode. Expect: the meter returns to the LCR Auto (or whichever) test screen; the calibration now applies to every subsequent reading until you calibrate again or the meter is reset.
4.3 · The guard line (TL-23)
"GUARD" is a separate terminal, wired to the meter's shield rather than one leg of the measurement. It provides a shield for the DUT, improvised test leads, or equipment sitting somewhere with high electrical interference. The manual's own note: test leads should be properly shielded to GUARD if necessary, especially at high or low impedance extremes where a small stray coupling is a large fraction of the signal.
4.4 · Decision matrix — when calibration is worth doing
| Situation | Calibrate? | Why |
|---|---|---|
| Measuring a range/frequency the accuracy tables mark with * (§11) | Yes, mandatory | The manual states these ranges only hold their printed accuracy after open/short cal |
| Very low impedance (small R, large C) or very high impedance (large R, small C, large L) | Yes | Fixture parasitics are a larger fraction of a small or extreme reading |
| Using an improvised or longer-than-stock test lead | Yes, mandatory | Longer leads add more series L and parallel C — exactly what cal removes; the manual warns against long improvised wires precisely because they add uncorrected error |
| Switching from TL-21 alligator leads to TL-22 SMD tweezers (or back) | Yes | Calibration is specific to the fixture in use — a different fixture means different parasitics |
| Mid-range component (kΩ, µF-to-nF, mH), same TL-21 leads you calibrated with recently | Optional | Parasitics are a small fraction of a mid-range reading; skip for a quick go/no-go check |
| Rapid SORTING-mode binning of a known-good part family, same session, same leads | Skippable | Pass/fail against a tolerance band is forgiving of small systematic offsets that cal would remove |
Sources: manual §II Open/short calibration and Equivalent circuit (pp.6–7); §IV.2 CAL (pp.15–17); §III Panel Illustration guard-line note (p.8); §VII Electrical Specification note 2–3 (p.35).
5 · LCR AUTO & FREQ Letting the meter guess, and telling it not to
The LCR AUTO key does two jobs depending on mode: in its default state it auto-detects whether your DUT is an inductor, a capacitor, or a resistor and measures accordingly; press it again and it steps through manual single-parameter modes instead. FREQ decides at what frequency the meter's test signal probes the part — and for a real, lossy component, that choice changes the answer.
5.1 · How auto-detect decides L vs C vs R
Power on, and the meter is already in Auto-LCR mode at 1 kHz — the factory default. Each press of LCR AUTO cycles: Auto-LCR → Auto-L → Auto-C → Auto-R → Auto-DCR → back to Auto-LCR. In Auto-LCR, the meter measures the impedance, then classifies it using the quality factor:
| Condition (as printed in the manual) | Mode selected | Secondary display |
|---|---|---|
|Q| < 0.2 | Auto-R | θ |
Q ≥ 0.2 | Auto-L | Q |
Q ≥ −0.2 | Auto-C | D |
Observed: bench-tested on a physical DE-5000, a 100 nF-nominal capacitor (measured 94.7 nF) wired in series with 1% metal-film resistors, sweeping Q across the ±0.2 boundary at 1 kHz (Xc ≈ 1681 Ω). An 8.2 kΩ resistor (D ≈ 4.88, Q ≈ −0.205) read Cs; a 10 kΩ resistor (D ≈ 5.95, Q ≈ −0.168) read Rs. The switch happens at |Q| ≈ 0.2 with Q negative — confirming the real rule is
Q ≤ −0.2 selects Auto-C, and the printed ≥ is a manual typo, not a quirk of this unit.
Once Auto-L or Auto-C is selected, ranging is fully automatic. The primary display shows the L, C, or R value; the secondary display shows the paired quality/dissipation figure. The manual groups these pairs explicitly: (L + Q), (C + D), (R + θ). One footnote exception: when LCR Auto is active and the DUT's capacitance is under 5 pF, the secondary display substitutes the equivalent parallel resistance Rp for D — D becomes numerically unstable at such a small capacitance.
Manual (single-parameter) modes work differently: press LCR AUTO to step to Auto-L, Auto-C, Auto-R, or Auto-DCR directly, and the meter measures only that parameter without trying to classify the DUT first. Auto-R mode and DCR mode carry no secondary display at all — there's no D/Q/θ concept for a pure resistance measurement in the manual's model.
5.2 · FREQ and why frequency changes the answer
Press FREQ to cycle the test signal frequency: 1 kHz → 10 kHz → 100 kHz → 100 Hz → 120 Hz → back to 1 kHz. Default is 1 kHz. The LCR impedance scale ranges are based on the selected test frequency — this isn't cosmetic, it changes what range the meter auto-selects internally.
For an ideal capacitor, C would read identically at every frequency. Real capacitors don't, because their dielectric losses and parasitic inductance are themselves frequency-dependent. The manual's own worked example, sweeping FREQ on the same physical capacitor in Auto-LCR mode:
| Frequency | Cs reading | D reading |
|---|---|---|
| 1 kHz | 666.8 nF | 0.004 |
| 10 kHz | 659.8 nF | 0.007 |
| 100 kHz | 646.6 nF | 0.009 |
| 100 Hz | 669.2 nF | 0.003 |
| 120 Hz | 669.2 nF | 0.004 |
The same physical part reads 666.8 nF at 1 kHz and 646.6 nF at 100 kHz — a 3% swing — while D climbs from 0.003 at 100 Hz to 0.009 at 100 kHz, more than doubling. Neither number is "wrong." Both are honest measurements of a real, lossy capacitor at different frequencies. This is precisely why a capacitance reading without a stated test frequency is an incomplete spec.
Sources: manual §IV.3 LCR AUTO (pp.18–19, incl. footnotes *1–*3); §IV.4 Inductance/Capacitance/Resistance Measurement (pp.20–21); §IV.5 FREQ (pp.22–23).
6 · D / Q / ESR / θ Four faces of the same loss, and which one to trust for what
D, Q, ESR, and θ are not four different measurements — they're four different ways of reporting the identical resistive-to-reactive ratio from §1's vector. The D/Q/ESR/θ key just changes which face the secondary display shows. Which one you should be looking at depends on what question you're asking.
6.1 · Definitions, straight from the manual
| Symbol | Name | Formula (as printed) |
|---|---|---|
| D | Dissipation factor | D = 1/Q = tanθ |
| Q | Quality factor | Q = Xs/Rs = 2πfLs/Rs = 1/(2πfCsRs) in series mode, or Q = B/G = Rp/|Xp| = Rp/(2πfLp) = 2πfCpRp in parallel mode |
| ESR | Equivalent series resistance | The series-mode resistive part, Rs, reported directly in ohms |
| θ | Phase angle | θ = tan⁻¹(Xs/Rs); positive for inductive, negative for capacitive DUTs |
Each press of D/Q/ESR/θ steps the secondary display: D → Q → ESR → θ → back to D. The manual's own worked example, all four readings taken on the identical DUT in Cs testing mode (1 kHz, Auto-LCR, primary reading held at Cs 666.8 nF throughout):
- From the manual's own example (§IV.9, p.27–28), not a live capture — no physical unit was tested for this guide.
- D 0.004 and Q 227 are reciprocal-consistent:
1/227 ≈ 0.0044, matching the printed D within rounding. - θ = −89.7° confirms the DUT reads capacitive (negative angle, per §1's sign convention) and close to a pure 90° — a low-loss capacitor.
- D/Q/ESR/θ has no function in LCR Auto mode — you must be in a manual L or C mode (or accept whichever the auto-classifier already picked) for this key to do anything, and it has no effect at all in Auto-R or DCR mode (no secondary parameter exists there).
6.2 · Sub-display accuracy — how errors propagate from Z to D, ESR, and θ
The manual defines Ae as the impedance (Z) accuracy — the number from the resistance/capacitance/inductance tables in §11 — and derives every secondary parameter's own accuracy from it:
Q = 1/D Rp = ESR × (1 + 1/D²)
D accuracy De = ± Ae × (1 + D)
ESR accuracy Re = ± ZM × Ae (Ω) where ZM = impedance calculated as 1/(2πfC) or 2πfL
Phase angle θe = ± (180/π) × Ae (deg)
The practical upshot: D and θ accuracy both degrade as the base impedance accuracy degrades — the same ranges marked * in §11's tables (worth calibrating for) also carry the least trustworthy secondary-parameter readings.
6.3 · Decision matrix — which secondary parameter to read
| Parameter | Best for | Read it when | Caution |
|---|---|---|---|
| D | Capacitors | Comparing against a datasheet D spec, or judging whether a cap has drifted from its rated loss | Meaningless without stating the test frequency (§5.2) |
| Q | Inductors | Comparing against a datasheet Q spec, or judging coil/core loss | Same frequency dependence as D; also unstable near Q≈0 (the R/L/C classification boundary in §5.1) |
| ESR | Capacitors in power/filtering roles (electrolytics, film caps in a supply) | You want an absolute ohms figure you can compare directly across parts, or check against a "how much did this cap age" rule of thumb | Only reported meaningfully in series-mode-dominant readings; not a resistor measurement (that's DCR/Auto-R) |
| θ | Quick eyeball of "how close to ideal", any L or C part | You just want to confirm the DUT is genuinely reactive (θ near ±90°) rather than mostly resistive | Least intuitive to read numerically; D or Q usually more actionable |
Sources: manual §II Measuring Principle Introduction, Q/D relationships (p.6); §IV.9 D/Q/ESR/θ (pp.27–28); §VII Electrical Specification, Sub-display parameters accuracy (p.36).
7 · SER/PAL Two descriptions of one impedance, and a 100 kΩ rule for choosing
§1 introduced series and parallel as two mathematically equivalent descriptions of the same impedance. This section is about the practical question: which description does the DE-5000 pick for you, and when should you override it.
7.1 · The two equivalent circuits
In series form, the DUT is modeled as a resistance and a reactance in series: Z = Rs + jXs. In parallel form, it's modeled as the reciprocal admittance Y = 1/Z = G + jB, with a parallel resistance Rp and parallel reactance Xp. Both circuits produce the identical impedance at a given frequency — they're algebraic duals, not two different physical models of the part.
7.2 · Why the choice matters: which loss mechanism dominates
The manual's reasoning, transcribed directly: real capacitors and inductors both have a series resistance Rs and a parallel resistance Rp in their equivalent circuit, but which one matters depends on the part's size.
| Part | If small | If large |
|---|---|---|
| Capacitor | Rp dominates → use parallel mode | Rs dominates → use series mode |
| Inductor | Rp has almost no effect → use series mode | Rs has no effect → use parallel mode |
Note the inductor and capacitor rules run in opposite directions: low-value inductor → series, low-value capacitor → parallel. This isn't arbitrary — it follows from which loss path (series lead resistance vs. parallel leakage) is physically larger relative to the reactance at that value.
7.3 · The 100 kΩ auto-threshold
SER/PAL has no function in LCR Auto mode — the meter picks for you, automatically, based on the total equivalent impedance it measures: above 100 kΩ, parallel mode is selected (Lp, Cp, or Rp shown); below 100 kΩ, series mode is selected (Ls, Cs, or Rs shown). In a manual (non-auto) L/C/R mode, press SER/PAL to force the opposite description at will. The manual's own toggling example, same DUT, same frequency, both readings 666.8 nF / D 0.004:
Cs testing mode → press SER/PAL → Cp testing mode → press SER/PAL → back to Cs
Observed: bench-tested on a physical DE-5000, bare 1% metal-film resistors (no reactive part) direct on the terminals in Auto-LCR: 1.49 kΩ → Rs; 4.69 kΩ → Rs; 7.49 kΩ → Rs; 8.25 kΩ → Rs; 9.94 kΩ → Rp; 33.05 kΩ → Rp. The real series↔parallel cutover for a resistive DUT falls between 8.25 kΩ and 9.94 kΩ — roughly an order of magnitude below the printed 100 kΩ, not a rounding difference. This was only characterized for the R family; whether C and L DUTs share the same real threshold, or the printed 100 kΩ, hasn't been separately bench-checked — Lab 2 below uses the manual's stated 100 kΩ for its arithmetic and that number should be treated as unverified for capacitance.
7.4 · Decision matrix — when to override the 100 kΩ auto-threshold
| Scenario | Force | Why |
|---|---|---|
| Small-value inductor near the 100 kΩ boundary (impedance borderline) | Series (Ls) | §7.2's rule: for a small inductor, Rp has almost no effect — series is the physically meaningful description regardless of which side of 100 kΩ the auto-threshold landed on |
| Large-value inductor (high reactance, comfortably >100 kΩ already) | Parallel (Lp) — usually already auto-selected | Rs has no effect at this size; auto-threshold and physical reasoning agree |
| Small-value / high-impedance capacitor (low pF range) | Parallel (Cp) — usually already auto-selected | Rp dominates; matches leakage-dominated real behavior |
| Large-value / low-impedance capacitor (µF electrolytic) | Series (Cs) — usually already auto-selected | Rs (ESR) dominates; matches how electrolytics actually fail |
| Matching how the part is actually wired in-circuit (e.g. an electrolytic in a series RC filter) | Whichever matches the circuit topology, even if it disagrees with the 100 kΩ rule | The auto-threshold is a heuristic about typical loss dominance, not a statement about your specific circuit |
Sources: manual §II Measuring Principle Introduction, Q/D discussion of Rs vs Rp dominance (p.6); §IV.10 SER/PAL (p.29).
8 · REL% & HOLD Compare against a reference, or freeze a reading in place
Neither key does anything in LCR Auto mode. Both are simple, single-purpose utilities once you're in a manual measurement mode: REL% turns every subsequent reading into a percentage deviation from a stored reference; HOLD just stops the display from updating.
8.1 · REL%
Press REL% to store the current reading as the reference value; the Δ icon appears, and the secondary display switches to showing the percentage deviation of every subsequent reading from that stored reference:
REL% = (DCUR − DREF) / DREF × 100%
where DCUR is the current DUT reading and DREF is the stored reference. The percentage range runs from −99.9% to 99.9%; if a new measurement is more than double the reference, the secondary display shows OL%. Press REL% for more than 2 seconds to exit relative mode.
The manual's own worked example (100 Hz, Auto-LCR, Cs mode) — quoted here, not bench-captured:
- From the manual's own example (§IV.11, pp.30–31), not a live capture.
- Check the arithmetic:
(194.16 − 669.3) / 669.3 × 100% ≈ −70.98%, matching the printed −70.9% within rounding. - REL% is built for comparing a batch of parts against one known-good reference — swap DUTs without leaving the mode and read the deviation directly instead of doing the subtraction yourself.
8.2 · HOLD
Press HOLD to freeze the primary display; the HOLD icon appears. In Hold mode, only the backlight key and the PC key retain function — everything else is locked out until you press HOLD again to release it.
Sources: manual §IV.11 REL (pp.30–31); §IV.12 HOLD (p.32).
9 · SORTING mode Turning a tolerance band into a PASS/FAIL buzzer
SORTING exists for one job: binning a stack of nominally-identical parts against a reference value and a tolerance, fast, without you doing the percent-difference math by hand. It's disabled in LCR Auto mode by design — you have to commit to a primary parameter first.
9.1 · Entering and configuring sorting mode
With a DUT connected and a manual L/C/R mode already selected, press SORTING to enter. The display automatically switches to 2000-digit resolution. If the reading is OL or below 200 counts, sorting is disabled — there isn't enough resolution to bin sensibly.
Inside sorting mode, the primary display shows PASS or FAIL based on whether the measured impedance falls inside the configured tolerance band; the secondary display keeps showing the raw measurement reading. Configure the reference value, range, and tolerance with SETUP, ◄/►, ▲/▼, and confirm with ENTER.
9.2 · The manual's own worked example
Starting point: Auto-LCR, 1 kHz, D 0.004, Cs 666.8 nF.
- Press SORTING. Expect: secondary display shows the reading, 667 nF; primary display shows
PASS(default Tol ±1%, 1 kHz). - Press SETUP. Expect: "Range" flashes; use ◄/► to pick the decimal-point position and units for the component you're sorting.
- The display shows
0667.with the last digit flashing. Expect: ◄/► selects which digit is active, ▲/▼ adjusts its value — set the reference value digit by digit, then press ENTER. - "Tol±" flashes. Expect: each press of ◄/► cycles the tolerance: ±1%, ±2%, ±5%, ±10%, ±20%, −20%/+80%, ±0.25%, ±0.5%.
- Press ENTER to confirm. Expect: setup is complete; the meter returns to live PASS/FAIL sorting against the configured reference and tolerance.
- Insert a DUT inside tolerance. Expect: the buzzer beeps once, primary display shows
PASS— the manual's own example readsCs 667 nF PASS. - Insert a DUT outside tolerance. Expect: primary display shows
FAIL— the manual's own example readsCs 673 nF FAIL.
From the manual's own example (§IV.7, pp.23–25), not a live capture — no physical unit was tested for this guide.
Sources: manual §IV.7 SORTING/SETUP/ENTER (pp.23–25); §III LCD Display Illustration, Sorting/Tol/Range icons (pp.11–12).
10 · PC / data logging One paragraph, because you don't own the adapter
The PC key starts and stops data transmission to a PC through the optional, fully isolated IR-to-USB interface — a separate accessory this guide's reader doesn't own, so it gets a paragraph rather than a module.
With the optional IR-to-USB case snapped onto the meter's rear IR slot (item 19 in §3.2) and connected by USB to a computer, pressing PC starts transmission and the PC icon appears on the display (icon 4 in §3.3); pressing PC again cancels it. The General Specification table (§11.1) lists the IR-to-USB case as an option, not a standard accessory — if you later add one, the manual's §IV.8 (p.26) is the four-step reference: snap on the case, connect USB, press PC, press PC again to stop.
Sources: manual §IV.8 PC (p.26); §VI General Specification, Option row (p.34).
11 · Accuracy & specification reference The numbers, transcribed exactly as printed
This section is a lookup table, not a narrative. Every figure below is copied from the manual's General Specification and Electrical Specification tables — treat it as the reference you return to, not something to read start-to-finish.
11.1 · General specification
| Item | Spec |
|---|---|
| Item | Dual Display L C R Meter |
| Parameters measured | Ls / Lp / Cs / Cp / Rs / Rp / D / Q / θ / ESR; LCR auto selection |
| Measuring circuit mode | Series / Parallel |
| Displays | Dual 19999 / 9999 display |
| Ranging mode | Auto |
| Measuring terminals | 4-wire sockets (terminals) & guard |
| Auto LCR test range | L: 20.000 µH ~ 2.000 KH · C: 200.00 pF ~ 20.00 mF · R: 20.000 Ω ~ 200.0 MΩ · DCR: 200.00 Ω ~ 200.0 MΩ |
| Test frequency | 100 Hz / 120 Hz / 1 kHz / 10 kHz / 100 kHz |
| Backlight | Yes |
| Tolerance mode | ±0.25%, ±0.5%, ±1%, ±2%, ±5%, ±10%, ±20%, −20%/+80% |
| Test signal level | 0.5 Vrms typ. |
| Measuring rate | 1.2 / second, nominal |
| Response time | Approx. 1 second / DUT |
| Auto power-off | Approx. 5 minutes without key operation (battery power only, §3.1) |
| Temperature coefficient | 0.15 × (spec. accur.) per °C (0–18°C, 28–50°C) |
| Operation temperature | 0°C to 50°C; 0–70% R.H. |
| Storage temperature | −20°C to +60°C; 0–80% R.H. |
| Battery voltage indication | Full icon when good; low-battery icon means replace immediately to stay within spec |
| Standard accessories | Alligator test lead case (TL-21) · AC/DC adaptor · guard line (TL-23) · user manual · DC 9 V battery |
| Option | IR to USB case · SMD tweezers (TL-22) |
| Dimensions (L/W/H) | 188 / 95 / 52.5 mm |
| Weight | Approx. 350 g (LCR meter only, excluding battery) |
±(% of reading + number of least-significant digits) at 23°C ±5°C, <75% R.H. A cell reading 0.3%+2 means: take 0.3% of the displayed value, then add 2 counts in the last displayed digit. Ranges marked * require open/short calibration (§4) to hit the printed number.
11.2 · Resistance accuracy (parallel / series mode)
| Range | Resolution | 100/120 Hz | 1 kHz | 10 kHz | 100 kHz |
|---|---|---|---|---|---|
| 20.000 Ω | 0.001 Ω | — | 1.0%+3* | 0.3%+2 | 0.3%+2 |
| 200.00 Ω | 0.01 Ω | — | 1.0%+3* | 0.3%+2 | 0.3%+2 |
| 2.0000 kΩ | 0.0001 kΩ | 0.3%+2 | 0.3%+2 | 0.3%+2 | 0.6%+3 |
| 20.000 kΩ | 0.001 kΩ | 0.3%+2 | 0.3%+2 | 0.3%+2 | 0.6%+3 |
| 200.00 kΩ | 0.01 kΩ | 0.5%+2 | 0.5%+2 | 0.5%+2 | 1.0%+3 |
| 2.0000 MΩ (2.000 MΩ) | 0.0001 MΩ | 1.0%+3 | 1.0%+3 | — | 2.0%+3* |
| 20.000 MΩ (20.00 MΩ) | 0.001 MΩ | 2.0%+3* | 2.0%+3* | — | — |
| 200.0 MΩ | 0.1 MΩ | 2.0%+3* | 2.0%+3* | — | — |
* Do open/short calibration before measuring on ranges marked with * to have better precision.
11.3 · DCR accuracy
| Range | Resolution | Accuracy |
|---|---|---|
| 200.00 Ω | 0.01 Ω | 1.0%+3* |
| 2.0000 kΩ | 0.0001 kΩ | 0.2%+2 |
| 20.000 kΩ | 0.001 kΩ | 0.2%+2 |
| 200.00 kΩ | 0.01 kΩ | 0.5%+2 |
| 2.0000 MΩ | 0.0001 MΩ | 1.0%+3 |
| 20.000 MΩ | 0.001 MΩ | 2.0%+3* |
| 200.0 MΩ | 0.1 MΩ | 2.0%+3* |
* Do open/short calibration before measuring on ranges marked with * to have better precision.
11.4 · Capacitance accuracy (parallel / series mode)
| Range | Resolution | 100/120 Hz | 1 kHz | 10 kHz | 100 kHz |
|---|---|---|---|---|---|
| 200.00 pF | 0.01 pF | — | — | 1.2%+5* | 2.0%+5* |
| 2000.0 pF | 0.1 pF | — | 2.0%+3* | 0.3%+2 | 0.6%+3 |
| 20.000 nF | 0.001 nF | 2.0%+3* | 0.3%+2 | 0.3%+2 | 0.6%+3 |
| 200.00 nF | 0.01 nF | 0.3%+2 | 0.3%+2 | 0.6%+2 | 1.2%+5* |
| 2000.0 nF | 0.1 nF | 0.3%+2 | 0.6%+2 | 2.0%+5* | — |
| 20.000 µF (20.00 µF) | 0.001 µF | — | 0.6%+2 | — | — |
| 200.00 µF (200.0 µF) | 0.01 µF | 0.6%+2 | 1.0%+3* | — | — |
| 2000.0 µF (2000 µF) | 1 µF | 1.0%+3* | — | 3.0%+5 (100 µF max.)* | — |
| 20.00 mF | 0.01 mF | 1.2%+3* | — | — | — |
If the reading is below 2000, the unit on display is pF (below 2000 counts, the meter shows the finer unit). * Do open/short calibration before measuring on ranges marked with * to have better precision.
11.5 · Inductance accuracy (parallel / series mode)
| Range | Resolution | 100/120 Hz | 1 kHz | 10 kHz | 100 kHz |
|---|---|---|---|---|---|
| 20.000 µH | 0.001 µH | — | — | — | 2.5%+5* |
| 200.00 µH | 0.01 µH | — | — | 1.2%+5* | 0.6%+3 |
| 2000.0 µH | 0.1 µH | — | 2.0%+5* | 0.6%+3 | 0.6%+3 |
| 20.000 mH | 0.001 mH | 1.2%+5* | 1.0%+5 | 0.3%+2 | 0.6%+3 |
| 200.00 mH | 0.01 mH | 0.3%+2 | 0.6%+2 | 0.3%+2 | 1.2%+5* |
| 2000.0 mH | 0.1 mH | 0.3%+2 | 0.3%+2 | 0.6%+3 | — |
| 20.000 H | 0.001 H | 0.3%+2 | 0.3%+2 | 0.6%+3 | — |
| 200.0 H | 0.1 H | 0.6%+3 | 1.2%+5* | — | — |
| 2.000 KH | 0.001 KH | 1.2%+5* | — | — | — |
If the reading is below 2000, the unit on display is µH. * Do open/short calibration before measuring on ranges marked with * to have better precision.
11.6 · Accuracy vs. resistance (ZDUT)
| ZDUT | DCR | 100/120 Hz | 1 kHz | 10 kHz | 100 kHz |
|---|---|---|---|---|---|
| 0.1 ~ 1 Ω | 1.2%+5* | 1.2%+5* | 1.2%+5* | 1.2%+5* | 2.5%+5* |
| 1 ~ 10 Ω | 0.6%+3* | 0.6%+3* | 0.6%+3* | 0.6%+3* | 1.2%+5* |
| 10 ~ 100 kΩ | 0.3%+2 | 0.3%+2 | 0.3%+2 | 0.3%+2 | 0.6%+3 |
| 100 k ~ 1 MΩ | 0.6%+3 | 0.6%+3 | 0.6%+3 | 0.6%+3 | 2.5%+5* |
| 1 M ~ 20 MΩ | 1.2%+5* | 1.2%+5* | 1.2%+5* | 2.5%+5* | 100 k~2 MΩ |
| >20 MΩ | 2.5%+5* | 2.5%+5* | 2.5%+5* | — | — |
* Do open/short calibration before measuring on ranges marked with * to have better precision.
Two correction rules apply on top of this table:
- If
D > 0.1, multiply the accuracy by√(1 + D²). - If
D ≪ 0.1: in capacitance mode,ZC = 1/(2πfC); in inductance mode,ZL = 2πfL— use the relevant Z to look up the row above.
Sources: manual §VI General Specification (p.34); §VII Electrical Specification, all tables (pp.35–36).
12 · Labs Run these on your own bench — this guide gives you the manual's numbers to check against
No physical DE-5000 was tested to write this guide, so these labs are structured around the manual's own worked examples rather than a live transcript. Follow each recipe on your own unit; the "Expect" lines quote what the manual documents happening on its unit for the identical sequence, not a capture from this session.
- Power on, connect TL-21 to the meter's terminals, leave the far end open (nothing touching, nothing shorted). Expect: default Auto-LCR mode, 1 kHz, per manual §IV.1.
- Hold CAL. Expect (manual, p.16): display shows
OPEnwith a blinking----. - Press CAL to start the open calibration. Expect (manual, p.16): a 30-second countdown (
OPEn -30-), thenOPEn PASS. - Press CAL, then short the TL-21 leads together. Expect (manual, p.17): display shows
Srt ----. - Press CAL to start the short calibration. Expect (manual, p.17): a second 30-second countdown (
Srt -30-), thenSrt PASS. - Press CAL once more to store both to EEPROM. Expect: the meter returns to its measurement screen; both parasitic corrections now apply.
FAIL instead of PASS, you almost certainly have something connected that shouldn't be (open step) or an incomplete short (short step). Re-verify the lead position and redo the whole sequence from open — the manual gives no partial-retry path.SER/PAL's 100 kΩ auto-threshold (§7.3) is a statement about impedance, not capacitance directly — and a capacitor's impedance at a given frequency is Xc = 1/(2πfC). That means the same physical part can sit on either side of the threshold depending purely on which FREQ you select. This is arithmetic from the manual's own formula (§1), not a live reading — work it out before you touch the meter, then confirm the mode annunciator matches your prediction.
- Pick a capacitor and compute the frequency at which its impedance crosses 100 kΩ:
f = 1/(2π × 100,000 × C). For a 666.8 nF part (the manual's own example value), that'sf ≈ 1/(2π × 100,000 × 666.8×10⁻⁹) ≈ 2.39 Hz— far below any of the DE-5000's five test frequencies, so a 667 nF cap will read in series mode (Cs) at every available FREQ setting. - Pick a much smaller part instead — say, a 100 pF capacitor:
f ≈ 1/(2π × 100,000 × 100×10⁻¹²) ≈ 15.9 MHz, also far above every available FREQ, so a 100 pF cap reads in parallel mode (Cp) at every available FREQ setting too. - Connect your actual part in Auto-LCR mode at 1 kHz. Expect: the LCD's Cs or Cp annunciator (icon 11, §3.3) tells you which side of 100 kΩ your part's impedance landed on at this frequency.
- Step through FREQ and watch the annunciator. Expect (inferred from the impedance formula, not manual-quoted): for a part whose impedance is genuinely close to 100 kΩ at some available frequency, the Cs/Cp annunciator flips as you cross it — most everyday nF-to-µF parts will stay on the Cs side at every available frequency, which is itself the useful observation.
- In a manual (non-Auto) mode, press SER/PAL to force the opposite description regardless of the auto-threshold. Expect (manual, p.29): the primary reading changes — Cs and Cp are not numerically identical for a real, lossy part (§7.4's "things that bite" callout).
Xc versus 100 kΩ, not just by reading the icon.- In Auto-LCR at 1 kHz, take a baseline reading of your reference part. Expect: a stable Cs (or Ls/Rs) reading with a D/Q value that isn't jumping around — noisy contact gives false FAILs later.
- Press SORTING. Expect (manual, p.24): secondary display shows the reading, primary shows
PASSorFAILagainst the default tolerance. - Press SETUP, set the reference value to your baseline reading digit-by-digit with ◄/► and ▲/▼, then ENTER. Expect (manual, p.24): "Tol±" flashes next.
- Pick a tolerance appropriate to your part's spec (e.g. ±5% for a 5%-tolerance resistor) with ◄/►, confirm with ENTER. Expect (manual, p.25): setup completes and the meter returns to live sorting.
- Reinsert the same part. Expect: buzzer beeps once,
PASS— matching the manual's own example,Cs 667 nF PASS. - Swap in a part you know is out of spec (or physically stress/flex a marginal one). Expect:
FAIL— matching the manual's own example,Cs 673 nF FAILfor a ±1% tolerance band around 667 nF.
Sources: manual §IV.2 CAL (pp.15–17); §II impedance formula (p.5); §IV.10 SER/PAL (p.29); §IV.7 SORTING (pp.23–25).
13 · Capstone One unknown part, every skill in this guide
Pull an unlabeled or unknown-value part from your junk box — ideally one you genuinely don't already know the value of — and take it through the full workflow. This is a checklist for your own bench, not a scripted transcript; the point is that you make each decision deliberately rather than accepting whatever the meter defaults to.
- Calibrate. Connect TL-21, run open then short calibration (§4.2, Lab 1). Confirm both steps report
PASSbefore continuing. - Measure blind, in Auto-LCR at 1 kHz. Connect the unknown part. Read the LCD icon row first (§3.3) — is
LCRplusLs/Lp,Cs/Cp, orRs/Rplit? That's your L/C/R identification, made by the meter's own|Q|classification logic (§5.1), not by you guessing from the part's appearance. - Read the rough value. Note the primary display's magnitude and unit. Sanity-check it against any markings on the part (color bands, printed code) if present — and note explicitly if there are none, which is presumably why it's unknown.
- Choose series or parallel deliberately. Note which the auto-threshold picked (§7.3). Using §7.2's Rs-vs-Rp reasoning and §7.4's decision matrix, decide whether that's also the physically meaningful description for a part of this apparent value — and if not, force the other with SER/PAL in a manual mode and note how much the primary reading moves.
- Read D/Q/ESR/θ appropriately. Using §6.3's decision matrix, pick the secondary parameter that matches what you'd actually want to know about this part (its loss, if it's a cap or inductor; its phase, if you're just sanity-checking "is this reactive at all"). State which one you picked and why.
- Decide pass/fail via SORTING. If you have a spec or a second, known-good sample of the same nominal part, configure SORTING (§9) with an appropriate tolerance and confirm your unknown part against it. If you don't have a reference, state what tolerance you'd choose and why, based on the part's apparent type and the accuracy tables in §11.
Sources: synthesizes §3–§9 of this guide, all traced to manual §III–§IV (pp.8–31).
14 · Troubleshooting Symptom → cause → fix
| Symptom | Likely cause | Fix |
|---|---|---|
CAL step shows FAIL | A lead or DUT was still connected during open or short calibration, adding uncorrected impedance | Remove everything for the open step; fully short the leads for the short step. Redo the whole open-then-short sequence from the start (§4.2) |
Primary display shows OL. | Reading is out of range for the current mode/range — could be a genuinely open circuit, a bad connection, or a part outside the Auto LCR test range (§11.1) | Check the lead connection first; if solid, confirm the part's expected value is inside the meter's 20.000 µH~2.000 KH / 200.00 pF~20.00 mF / 20.000 Ω~200.0 MΩ range |
| SORTING key does nothing | You're still in LCR Auto mode — SORTING (like SER/PAL, REL%, D/Q/ESR/θ) requires a manual primary-parameter mode first | Press LCR AUTO to step into Auto-L, Auto-C, Auto-R or Auto-DCR, then try again |
| SORTING won't engage even in a manual mode | Reading is OL, or below 200 counts — not enough resolution to bin | Confirm a stable, in-range reading first; sorting needs the meter to already trust its own number |
| Battery icon shows one bar / low | 9 V battery is depleted | Replace immediately — the manual explicitly ties continued spec accuracy to prompt replacement (§3.1, §11.1) |
| Meter beeps twice on a keypress and nothing happens | Not a fault — that key has no function in the current mode (e.g. SER/PAL in LCR Auto mode) | Check which mode you're in; a single beep means the key worked, a double beep means it was a no-op by design |
APO icon never appears, meter never auto-powers-off | Running from the AC/DC adaptor — auto power-off disables itself automatically on adaptor power | Expected behavior, not a fault. Switch to battery power if you want APO active |
| Same part reads differently after switching leads (TL-21 ↔ TL-22) | Calibration is fixture-specific; a new fixture carries the old fixture's stored correction | Re-run open/short calibration with the new fixture attached (§4.4 decision matrix) |
| Reading jumps around / D or Q looks unstable | Poor contact, a DUT that's still charged (for capacitors), or measuring right at a range boundary | Discharge the capacitor before testing (manual warning, p.2 and p.21); check lead contact; if near a range boundary, small ambient noise can flip the last digit |
| Toggling SER/PAL changes the primary reading | Not a fault — Cs and Cp (or Ls and Lp) are only identical for an ideal, lossless part (§7.4 "things that bite") | Expected for any real component with nonzero D; pick the mode that matches how the part is actually used (§7.4) |
Sources: manual general warnings (p.2); §IV.1 Power ON/OFF (pp.13–14); §IV.2 CAL (pp.15–17); §IV.7 SORTING (pp.23–25); §V Replacing Batteries (p.33).
15 · Cheat sheet One page to tape to the bench
Everything you need at the terminal block, with the section that explains it in brackets.
Universal starting pattern (§4.2)
- Connect TL-21. Hold CAL → OPEN (leads open) → SHORT (leads shorted) → both PASS → CAL to store.
- Connect DUT. Default is Auto-LCR, 1 kHz.
- Read the icon row first (LCR / Ls·Lp / Cs·Cp / Rs·Rp), then the digits.
Key grammar (§3.2)
- LCR AUTO cycles Auto-LCR → Auto-L → Auto-C → Auto-R → Auto-DCR
- FREQ cycles 1 kHz → 10 kHz → 100 kHz → 100 Hz → 120 Hz
- D/Q/ESR/θ · SER/PAL · REL% · SORTING — all dead in LCR Auto mode
- CAL enters calibration · REL% exits relative mode
Auto L/C/R classification (§5.1)
|Q| < 0.2→ Auto-R, secondary θQ ≥ 0.2→ Auto-L, secondary QQ ≥ −0.2(printed; likely means ≤ −0.2) → Auto-C, secondary D- C < 5 pF in LCR Auto → secondary shows Rp instead of D
Series vs parallel (§7)
- Auto: >100 kΩ → parallel, <100 kΩ → series
- Small cap → parallel · large cap → series
- Small inductor → series · large inductor → parallel
- Override with SER/PAL in a manual mode
D/Q/ESR/θ quick pick (§6.3)
- D — capacitors, datasheet comparisons
- Q — inductors, datasheet comparisons
- ESR — capacitor absolute ohms, power/filter roles
- θ — quick "how reactive is this really"
Tolerance bins (§9)
- ±0.25% · ±0.5% · ±1% · ±2% · ±5% · ±10% · ±20% · −20%/+80%
- Disabled below 200 counts or on
OL
16 · Glossary Every term of art, with where it was taught
- Admittance (Y)
- The reciprocal of impedance,
Y = 1/Z = G + jB: the parallel-mode dual of impedance. (§1) - APO
- Auto power-off: shuts the meter down after ~5 minutes idle, but only on battery power — the AC/DC adaptor disables it. (§3.1)
- Conductance (G)
- The real (resistive) part of admittance, the parallel-mode counterpart to resistance. (§1)
- D (dissipation factor)
D = 1/Q = tanθ; how lossy a capacitor is relative to ideal. (§6.1)- DCR
- DC resistance mode; no secondary parameter available. (§5.1)
- DUT
- Device under test — whatever is connected to the terminals. (§1)
- ESR
- Equivalent series resistance — the series-mode resistive part
Rs, reported directly in ohms. (§6.1) - Fixture parasitic impedance
- The small series impedance and parallel admittance every test lead/socket adds, corrected out by open/short calibration. (§4.1)
- GUARD
- A shield terminal (used with the TL-23 guard line) that protects the DUT or improvised leads from interference, separate from the ± measurement path. (§4.3)
- HOLD
- Freezes the primary display; only the backlight and PC keys keep working while active. (§8.2)
- Impedance (Z)
- The complex quantity
Z = Rs + jXsthat every DE-5000 reading is ultimately derived from. (§1) - LCR AUTO
- The primary-parameter key; in its default state, auto-classifies the DUT as L, C, or R using
|Q|thresholds. (§5.1) - OL
- Overload/out-of-range indication; also disables SORTING mode. (§14)
- Open calibration
- Measures the fixture's parallel admittance with nothing connected, shown as
OPEnon the display. (§4.2) - Q (quality factor)
Q = Xs/Rsin series mode orQ = Rp/|Xp|in parallel mode; how close to ideal an inductor is. (§6.1)- REL%
- Relative mode: stores a reference reading and shows every later reading as a percentage deviation from it. (§8.1)
- Reactance (X)
- The imaginary part of impedance; inductive
XL = 2πfL(positive θ) or capacitiveXC = 1/(2πfC)(negative θ). (§1) - Series / parallel mode
- Two algebraically equivalent descriptions of the same impedance — Rs/Xs in series, Rp/Xp (or G/B) in parallel. (§7.1)
- Short calibration
- Measures the fixture's series impedance with the leads shorted, shown as
Srton the display. (§4.2) - SORTING
- Tolerance-binning mode: configures a reference value and tolerance, reports PASS/FAIL against it. (§9)
- Susceptance (B)
- The imaginary (reactive) part of admittance, the parallel-mode counterpart to reactance. (§1)
- θ (phase angle)
θ = tan⁻¹(Xs/Rs); positive for inductive DUTs, negative for capacitive ones. (§1, §6.1)- TL-21
- The standard alligator test lead case, used throughout this guide's labs. (§3.2)
- TL-22
- Optional SMD tweezers accessory, not owned by this guide's reference reader. (§3.2)
- TL-23
- The guard line accessory; shields the DUT or improvised leads from interference. (§4.3)
- Tolerance (sorting)
- The PASS/FAIL band configured in SORTING mode: ±0.25% through ±20%, or −20%/+80%. (§9.1)
- ZDUT
- The true DUT impedance the meter solves for once open and short calibration are both stored. (§4.1)
17 · Index Terms, keys, modes, and settings
- Admittance (Y)
- APO (auto power-off)
- Auto-DCR mode
- Auto-detect (|Q| thresholds)
- Accessories (TL-21/22/23)
- Accuracy notation (%+digits)
18 · Quiz Sixteen flashcards, self-scored
Reveal, then be honest. Aim for 13 of 16 before you trust a calibrated reading on a part you can't replace. Your score persists in this browser; the reset button below clears it.
Y_OPEN = Go + jωCo (nothing connected). Short isolates the fixture's series impedance Z_SHORT = Rs + jωLs (leads shorted). (§4.1)Q ≤ −0.2 — a real DE-5000 switched to Cs only once Q dropped past −0.2, not the other way around. (§5.1)D = 1/Q = tanθ. (§6.1)REL% = (DCUR − DREF) / DREF × 100%. OL% appears when a new reading is more than double the stored reference. (§8.1)0.3%+2 mean when applied to a displayed reading?19 · Sources Where every claim came from
Written against the vendor's DE-5000 Instruction Manual — Dual Display L C R Meter (document code HCA500000-00020, 19 printed pages, DER EE Electrical Instrument Co., Ltd.), read in full at /Volumes/thunderbay-public/documentation/DE-5000_manual_english.pdf on 2026-09-11. Every worked example, LCD screen quote, and spec-table figure is transcribed directly from the manual and is labelled as coming from the manual, never presented as a live capture. Where the manual's own text appeared internally inconsistent (the Auto-C threshold footnote in §5.1, the SER/PAL 100 kΩ threshold in §7.3), this guide quotes the printed text exactly, then reports what a physical DE-5000 actually did against real 1%-tolerance resistors and a measured capacitor — see the "Documented vs. observed" callouts in §5.1 and §7.3, dated 2026-09-13. Those two spot-checks are the only bench measurements behind this guide; no other section's figures were re-verified against a live unit. Formulas applied outside the manual's own worked numbers (the frequency-crossing arithmetic in Lab 2) are ordinary algebra on the manual's own equations, clearly marked as calculated rather than measured. External links below require a network; this guide itself works entirely offline.
- DE-5000 Instruction Manual (PDF, DER EE Electrical Instrument Co., Ltd., document HCA500000-00020). The sole primary source for this guide: measuring-principle theory, panel and LCD illustrations, every operation-instruction section, battery replacement, and both specification tables. Best used for the exact figures this guide's tables and diagrams were redrawn from.
- www.deree.com.tw — the vendor site printed in the manual's closing page. Domain as printed in the manual; not crawled or verified for this guide.
- Electrical impedance (Wikipedia) — general background on the Z = R + jX model behind §1; not manual-specific.
- Dissipation factor (Wikipedia) — general background on D as a capacitor loss metric, behind §6.
- Q factor (Wikipedia) — general background on Q as an inductor/resonator figure of merit, behind §6.
- Series and parallel circuits (Wikipedia) — general background behind §7's duality discussion.
- LCR meter (Wikipedia) — general background on how bridge-based LCR meters work, for context beyond this one model.
A guide built entirely from a manual is a transcript, not a bench test. Numbers in §11's accuracy tables, the worked examples in §5, §6, §8, and §9, and every "Expect" line in §12's labs describe what the manual documents — confirm them against the screen in front of you before trusting a measurement that matters.